Novasinergia vol.2 no.2 Riobamba jun./nov. 2019  Epub 10-Dic-2019

https://doi.org/10.37135/unach.ns.001.04.08 

Research Article

Detección de Cd (II) y Pb (II) por voltamperometría de separación anódica utilizando electrodos de carbono vidriosos modificados con aleación bimetálica Ag-Hg y Ag-Bi

Detection of Cd (II) and Pb (II) by anodic stripping voltammetry using glassy carbon electrodes modified with Ag-Hg and Ag-Bi bimetallic alloyed

Danny Valera1 
http://orcid.org/0000-0002-5977-329X

Mireya Sánchez2 
http://orcid.org/0000-0003-1762-9502

José Domínguez1 
http://orcid.org/0000-0003-1504-8712

Patricio J. Espinoza-Montero2 
http://orcid.org/0000-0003-0592-8652

Carlos Velasco-Medina3 
http://orcid.org/0000-0002-4488-4122

Patricio Carrera4 

Lenys Fernández1  2  * 
http://orcid.org/0000-0001-6720-6343

1 Departamento de Química, Universidad Simón Bolívar, Caracas, Venezuela, 89000

2 Escuela de Ciencias Químicas, Pontificia Universidad Católica del Ecuador, Quito, Ecuador, 170525

3 Facultad de Ingeniería Química, Escuela Politécnica Nacional, Quito, Ecuador, 170525

4Hidroecuador, Quito, Ecuador, 170801; dannyvalera@usb.ve;mires-40@hotmail.com; patoespinozamon@hotmail.com; carlos.velasco@epn.edu.ec; info@nanoinstrumentos.net


Resumen:

Presentamos la evaluación de los electrodos de carbono vítreo (GC) modificados con películas de Nafion (Nf) y depósitos bimetálicos de aleaciones Ag-Hg y Ag-Bi. La mayoría de los bimetálicos depositados, con un tamaño promedio de aproximadamente 150 nm, se dispersaron e incrustaron uniformemente dentro de la red del Nafion. En contraste, una cantidad mucho menor permaneció encima de las incrustadas sin ninguna orientación regular, como lo indica imágenes de microscopía electrónica de barrido (SEM) y microscopía de fuerza atómica (AFM). Las pruebas de sensibilidad para el electrodo modificado AgBiNf/GC produjeron límites de detección (DL), basados en la variabilidad de una solución en blanco (criterio de 3 s), de 0,78 y 0,66 μg L-1 para Cd (II) y Pb (II), respectivamente; mientras que el DL sobre el electrodo modificado AgHgNf/GC fue 0,17 y 0,24 μg L-1 para Cd (II) y Pb (II), respectivamente. La precisión de la metodología por voltametría de redisolución anódica se verificó mediante el cálculo de porcentajes de recuperación de los dos analitos, expresados como error relativo. Se logró una recuperación del 99% al 92%.

Palabras clave: Aleación bimetálica; electrodos modificados, voltamperometría de redisolución anódica de Pb y Cd

Abstract:

We report the evaluation of glassy carbon (G.C.) electrodes modified with Nafion (Nf) films and bimetallic alloyed Ag-Hg and Ag-Bi deposits. Most of the deposited bimetallic, with an average size of approximately 150 nm, were uniformly dispersed and embedded inside the Nafion net cells. In contrast, a much lesser amount of them remained on top of the embedded ones without any regular orientation, as indicated by scanning electron microscopy (SEM) and atomic force microscopy (AFM) images. Sensitivity tests for the AgBiNf/G.C. modified electrode produced detection limits (DL), based on the variability of a blank solution (3 s criterion), of 0.78 and 0.66 μg L-1 for Cd(II) and Pb(II), respectively, whereas for the AgHgNf/G.C. modified electrode D.L.s were 0.17 and 0.24 μg L-1 for Cd(II) and Pb(II), respectively. Accuracy of the anodic re-dissolution voltammetry methodology was checked by calculation of percentages of recovery of the two analytes, expressed as a relative error. Recovery from 99 % to 92 % was achieved.

Keywords: Bimetallic alloyed; modified electrodes, Pb and Cd anodic stripping voltammetry

1 Introduction

Monitoring heavy metals in the environment and biological samples is of utmost importance for evaluating and preserving the environment and public health quality levels.

One of the best ways of performing this monitoring is through sensitive and fast response sensors used in situ. Sensors with these characteristics provide real-time results useful to establish reference locations needed to follow the contaminants' behavior as a function of location and time. This type of study is increasingly necessary because heavy metals concentration in the environment keeps growing due to the larger augment in global anthropogenic activities. Consequently, more and more heavy metals get into the environment and, through the food chain, into the human organism (Karri et al., 2016). Among the contaminants, Cd(II) and Pb(II) are the ones that could be most dangerous to human health (Wan et al., 2010).

Cadmium could get into the human organism via inhalation or by food or drink ingestion. It could act as a cancer promoter by altering the regular genetic expression, inhibiting DNA reconversion, and inducing oxidative stress (Singh et al., 2009; Lin et al., 2016). The lead could get into the human organism similarly, as Cd does. Lead, as a contaminant, among other effects, could supplant calcium in proteins causing protein malfunctioning and thus impairing physiological functions (Zhang et al., 2015). Since Cd and Pb appear not to have known natural biological functions, the presence of these elements in the human body, at any concentration level, must be due to contamination. Institutions such as the Center for Disease Control (CDC), International Agency for Cancer Research (IACR), and World Health Organization (WHO) have dictated maximum concentration limits permitted for these metals, including Cd and Pb, in food, drinks, biological and environmental samples (Gumpu et al.,2015). Cd and Pb concentration levels in non-contaminated geological and environmental samples are usually relatively low, 10 (g L-1. However, those low levels must be accurately known to start a study regarding the extent of a possible contamination process by these two elements as a function of time.

Evaluation of electrochemistry-based techniques together with the best way of preconcentration and isolation of heavy metals for their accurate determination have been the goal of several research studies (Zhu et al., 2006; Liang et al., 2005; Abkenar et al.,2010; Abkenar et al., 2011; Hosseini et al., 2010; Dahaghin et al., 2017a,b; Bagheri et al., 2012; Huang et al.,2014; Huang et al., 2011; Bhatluri et al., 2017). Among the most widely used electrochemistry techniques for Cd and Pb determination, based on their high sensitivity and freedom from matrix effects, anodic re-dissolution voltammetry (ASV) is one of the most convenient.

During the first step of the ASV methodology, the analyte's ions in solution are reduced by electrolysis under reducing potential conditions, by way of which the analyte is separated from most of the sample's matrix freeing its measurement from most matrix effects. The reduced analyte's preconcentration follows this step by its deposition on the working sensor's surface, which increases the detection capacity of the technique. Finally, the analyte is re-dissolved under anodic re-dissolution conditions originating an electrical signal proportional to the analyte's concentration.

Fabrication of an electrode, which could function close to the ideal one for lead and cadmium determination at trace level in complex matrices, has been the objective of many research studies (Xiong et al., 2016). In most cases, mercury and mercury film-based electrodes are selected due to their excellent re-dissolution capacity (Zhu et al., 2004; Lakshmi et al., 2007; Ferreira et al.,2002). However, the presence of a highly toxic element, as in massive mercury sensors, makes them inadequate from the point of view of contamination (Cargnelutti et al., 2006). To further diminishing toxicity, bismuth, a more ecologic element, was chosen instead of mercury as a modifier for the sensor (Pei et al., 2014; Borgo et al., 2015). Based on those findings, we have tried several glassy carbon (G.C.) modified electrodes for the determination of Cd(II) and Pb(II) in human serum and urine samples (Valera et al., 2018).

In the present work, we report on the determination of Cd and Pb in human serum and urine samples using two electrodes, the first one consisting of a G.C. electrode covered by a Nafion (Nf) film doped with HgAg alloyed and the second one by BiAg alloyed.

Experimental

2.1 Reagents

Nafion (5% (w/w)) was purchased from Aldrich; BiNO3.5H2O(98%), cadmium nitrate (99.5%); lead nitrate (99.5%), and hydrogen peroxide (6% (w/w)) were purchased from Merck; silver nitrate, 99.8%, nitric acid (65%), acetic acid (99.8%) and ethanol (99.8%), were purchased from Riedel- de Haën.; potassium hydroxide (87.8%) was purchased from J.T. Baker; sodium hydroxide (98%) was purchased from Eka Nobel; sodium acetate (98%), and dimethylformamide (DMF) were purchased from Sigma. All reagents used were analytical grade reagents or better unless otherwise stated. Aqueous solutions were prepared using distilled/deionized, 18M ( cm-1 Millipore water.

2.2 Instrumental

Electrochemical measurements were carried out using a Princeton Applied Research (PAR), 273A model Galvanostat/Potenciostat, computer-controlled by the 270/250 Research Electrochemistry Software 4.23 from PAR, coupled to a conventional 15 mL three-electrode reaction cell. The modified electrodes in the evaluation were used as working electrodes; Ag/AgCl electrodes were used as a reference, and 0.5 cm diameter platinum wires were the counter electrodes. Scanning electron microscopy (SEM) analyses were performed using a PHENOM PROX tabletop scanning electron microscope. A PARK SYSTEMS equipment (NX10) was used for the ex-situ atomic force microscopy (AFM) studies.

2.3 Deposition of alloyed bimetallic deposits

G.C. electrodes were chosen as substrates for the deposition of the alloyed bimetallic to produce the modified electrodes. As a previous step, the G.C. electrodes were firstly polished using the number 2000 sandpaper and afterward using spreads of aluminum oxide powder, with decreasing particle size 1.0 µm, 0.3 µm, and 0.05 µm, on a billiard tablecloth. After polishing, the G.C. electrode was submerged in an ultrasonic bath containing distilled/deionized water for 5 minutes to eliminate any aluminum oxide particle loosely adhered to the glassy carbon. Five microliters of a 1% Nf solution were cast on the G.C. electrode using a micropipette. Immediately, 3 µL of pure DMF was added with a micropipette. Ethanol and DMF used as solvents were evaporated by heating at 30oC with an air gun and rotating the electrode at 50 rpm.

After this treatment, the electrode, already covered with an Nf film, was submerged in an aqueous solution having either 85% Ag and 15% Hg or 85% Ag and 15% Bi for 3 hours. Afterward, the electrode was carefully washed with distilled/deionized water to remove non-absorbed material. Finally, the metallic ions trapped in the Nf net are reduced by submitting the electrode to Coulombimetry at controlled potential (CPC) at -1.2 V for 300 seconds in a 1mol L-1 KNO3+ 0.1 mol L-1 HNO3 solution.

2.4 Sample recollection and treatment

2.4.1 Blood

A modified method according (Valera et al., 2018). Professional nurses collected blood samples directly into new "Vacutainer" tubes containing EDTA as an anticoagulant. After gentle shaking to promote contact with de EDTA solution, the samples were refrigerated until taken for analysis. For analysis, blood samples were centrifuged at 3000 rpm for 15 minutes to separate serum from the rest of the sample. Two milliliters of serum were diluted with 15 mL distilled/deionized water under gentle shaking and 8 mL conc. Nitric acid was added, and the sample was left shaking for 5 minutes to initiate digestion. After this time, 1 mL of a 30 % hydrogen peroxide solution was added. The samples were ultrasonicated for 30 minutes at 60 °C, then diluted to 25 mL with distilled/deionized water, which renders them ready for analysis.

2.4.2 Urine

A modified method according (Valera et al., 2018). Urine samples were collected by donors in virgin urine collection vessels. Donors were informed about the study and instructed on the best way of avoiding contamination during collection. Samples were immediately acidified with 2 mL conc. Nitric acid to avoid metal hydrolysis and microorganism growth and refrigerated until taken for digestion. Digestion was initiated on 10 mL of filtered samples (Whatman N° 1" filter paper) by adding 1 mL conc. Nitric acid plus 1 mL of a 30% hydrogen peroxide solution. Samples were then left under gentle shaking for 5 minutes and ultrasonicated for 30 min. at 60°C. Finally, samples were diluted to 25 mL with distilled/deionized water rendering them ready for analysis.

2.5 Determination of Cd (II) and Pb (II) by Anodic Stripping Voltammetry

For the preconcentration step, a 15 mL of a 0.1 mol L-1, pH 4.5, acetate buffer solution, and precisely known volumes of Cd(II) and Pb(II) standard solutions were added to a three-electrode conventional cell. The cell was then purged with an argon flow (20 mL min-1) for 5 minutes, and a -1.2 V potential was applied for 120 s, with continuous agitation followed by a 10 s period with no agitation.

Differential pulse voltammetry (DPV) scanning at 20 mV s-1 within the range from -650 to -350 mV vs. Ag/AgCl, was used for the re-dissolution step. After each concentration/re-dissolution cycle, the electrode's Cleaning was achieved by applying a -200 V vs. Ag/AgCl potential for 15 s.

Results and Discussion

3.1 Modified electrodes characterization

Figure 1a shows a micrograph of the Nf film covering the G.C. electrode (Valera et al.,2018), which is the first step in preparing the modified Nf/G.C. electrode. The whole Nf film seems flat in the figure, presenting some holes, approximately 900 to 950 nm in diameter, seen as dark spots.

Figure 1b shows a micrograph of the resulting AgBiNf/G.C. electrode's surface, obtained after 180 min (Valera et al.,2018). Immersion of the Nf covered the G.C. electrode in a 0.85 µg L-1 Ag + 0.15 µg L-1 Bi solution. SEM images allow for appreciation of metallic particles, white spots inside the Nf film. Particles with an approximate particle diameter of 150 nm were obtained.

The images of the surfaces after individual immersion for 180 minutes of G.C. electrodes in an Ag-Hg solution to produce the AgHgNf/G.C. electrode are shown in figure 1c (Valera et al., 2018).

Although the particle size of the AgBi alloys seems to be smaller ((150 nm) than the particle size of the AgHg ((200 nm), the latter shows a more dispersed distribution. AFM images are shown in figure 2. These images suggest that the Nf film completely covers the G.C. electrode (figure 2a). After deposition of the bimetallic particles, the images suggest that the particles are three-dimensionally dispersed throughout the polymer film, figure 2b (Valera et al., 2018).

3.2 Nafion film thickness

The simplest way of covering the G.C. electrode with an Nf film is by applying a small volume of an Nf solution to the electrode followed by adding a small volume of DMF to increase the polymeric film's stability (Kefala et al., 2004). Nf film thickness can be controlled by adding to the glassy carbon substrate a given volume of solutions containing different Nf concentrations.

We added fixed volumes of solutions containing Nf concentrations of 1% (w/v). Assuming a uniform distribution of the Nf solution on the glassy carbon electrode's surface (as indicated by the SEM images), film thickness I Na can be calculated, to a first approximation, by the formula I Na = m Na / πR 2 d Na (Valera et al.,2018); where m Na is the mass of Nf deposited on the electrode; d Na is the density of the Nf film (1.58 g cm-3), and R is the radius of the glassy carbon electrode (1.5 mm). With the addition of 5 µL of solutions containing 1.0 % w/v Nf and a volume of DMF of 3 µL, films with thickness 4.47 µm were obtained.

Preconcentration potential

Starting with 120 s preconcentration time, the influence of Cd's preconcentration potential and Pb's re-dissolution were studied in a -0.8 V to -1.3 V potential range.

The results depicted in figure 3a indicate that current peaks increase from -0.9 V to -1.2 V, and peaks start decreasing at -1.2 V. Given these results, -1.2 V was chosen as the optimal re-dissolution potential for best preconcentration of Cd and Pb.

3.4 Pre-concentration time

As preconcentration time plays a significant role in analyte's accumulation, a 1.2 V potential was applied to a 50 µg L-1 de Cd(II) and Pb(II) solution to find out the amount of Cd and Pb accumulated on the AgHgNf/G.C. and the AgBiNf/G.C. electrodes as a function of time accumulation was estimated by measuring the redissolution charge (µC) for each metal (figure 3b). Increasing time from 0 to 300 s resulted in augmented re-dissolution currents, indicating increased amounts of each analyte on the electrode's surface. After 300 s, the current starts to stabilize and keeps its value up to 400 s and even longer times; 120 s was selected as preconcentration time for all measurements.

3.5 Working range and sensitivity

Figure 4a depicts the working curves for the simultaneous re-dissolution of Cd (blue dots) and Pb (orange dots), using the AgBiNf/G.C. electrode. Both analytes behave similarly regarding their current-concentration linear relationship. The lead response is seen to be more sensitive. Linearity is kept from around 10 to 120 µg L-1, being limited by electrode surface saturation. After a 120 s preconcentration time, detection limits calculated based on the 3s criterion measuring blank solutions were 0.78 µg L-1 for Cd(II) and 0.66 µg L-1 for Pb(II), n = 6. Results obtained using the AgHgNf/G.C. electrode are shown in figure 4b. This figure was constructed in the same way as figure 4a using the corresponding current values from the voltammograms shown in the inset. Both analytes respond quite differently at this electrode. Sensitivity for Pb is higher than for Cd; the linear range is the same for both metals, meaning that the process of electrode surface saturation is very similar for both electrodes. After a preconcentration time of 120 s detection limits calculated based on the 3 s criterion measuring blank solutions were 0.17 µg L-1 for Cd(II) and 0.24 µg.L-1 for Pb(II), n = 6

Figure 1: SEM images of the Nf film covering the G.C. electrode and the Nf covered G.C. electrode surface after (a) 60 s. (b) 180 minutes of immersion of the Nf covered G.C. electrode in a 0.85 mg L-1 Ag + 0.15 mg L -1 Bi solution and then reducing the metallic ions trapped in the Nf net by CPC, and (c) After 180 minutes of immersion of the Nf covered G.C. electrode in a 0.85 mg L -1 Ag + 0.15 mg L -1 Hg solution and then reducing the metallic ions trapped in the Nf net by CPC (Valera et al., 2018

Figure 2: AFM images: (a) High magnification images (scan region 5 mm), showing the thickness of the AgBi nanoparticle. (b) AgBiNpNf on the G.C. substrate; the image shows the three-dimensional arrangement of alloyed nanoparticles, Ag-Bi (Valera et al., 2018). 

Figure 3: (a) Effect of the analyte preconcentration potential on the charge. (b) Effect of analyte preconcentration time on determination of 50 mgL-1 Pb (II). 

Figure 4: (a) Calibration curves at AgHgNf / CV electrode. (b) Calibration curves at AgBiNf / CV electrode. 

Figure 5: Repeatability study for measurements of Pb(II) and Cd(II) by ASV using the AgBiNf/G.C. and AgHgNf/G.C. electrodes. 

3.6 Precision of measurements

After a 120 s preconcentration time from a 40 µg L-1 Cd and Pb solution, twenty consecutive determinations of each metal were carried out, figure 5. Relative standard deviation (RSD) values range from 1.18 to 4.61. These could be considered outstanding RSD values attesting to an exact methodology.

3.7 Determination of Cd and Pb in human blood serum and urine samples

Modified electrodes were applied to Cd and Pb's determination in human blood serum and urine samples collected from donors living in a rural area that was part of a broader environment pollution study. Samples were collected from students of the University (Universidad Simon Bolivar, Caracas-Venezuela), doped with 40 µg L-1 of both Cd(II), and Pb(II), and analyzed following the analytical protocol described in section 2.5.

The ASV methodology's accuracy was checked by calculation of percentages of recovery of the two analytes, expressed as a relative error, R (table 1).

Results in table 1 demonstrate that Pb's determination can be achieved, with recoveries within 96 to 99 % in both types of samples, serum, and urine, using any new electrodes.

This accuracy indicates that these samples' organic matrices do not represent a limitation in Pb's determination. However, this is not the case for the determination of Cd in blood serum samples for which R% was 80% using the AgBiNf/G.C. electrode and even more extensive, 53%, using the AgHgNf/G.C. electrode.

These results show that AgBi alloyed are more convenient for modifying Nafion covered G.C. electrodes destined for analysis of organic samples with complex matrices than the AgHg alloy. Modifying the Nf/G.C. electrodes with AgHg alloyed nanoparticles produces electrodes unsuitable for Cd analysis in blood serum due to their unacceptable inaccuracy. Nonetheless, for urine, a sample with a lighter matrix than blood serum, the AgHgNf/G.C. electrode produces accurate results with recoveries within 92 to 94%.

Table 1: Results 

Analyte AgBiNf/GC found (µgL-1) AgHgNf/GC found (µg L-1) AgBiNf/GC found (µg L-1) AgHgNf/GC found (µg L-1) AgBiNf/GC R(%) AgHgNf/GC R (%)
Serum Serum Urine Urine Serum Urine Serum Urine
Cd (II) 41.16± 4.12 29.35 ± 4.43 46.23 ± 2.77 46.79 ± 2.66 80.32 92.46 52.84 93.59
Pb (II) 48.27 ± 1.94 48.69 ± 1.91 49.01± 2.15 49.48 ± 2.92 96.54 98.02 97.37 98.96

Conclusion

Results show that the new electrodes could be useful for Pb determination in human serum and urine samples with acceptable accuracy and precision. The new electrodes' ASV technique offers a faster, lower cost, more mobile, and easier to operate alternative. The AgHgNf/G.C. electrode has the advantage of being slightly more sensitive than the AgBiNf/G.C. for Cd and Pb determination, but it is not useful for Cd determination in serum samples due to lack of accuracy. Even though the AgHgNf/G.C. electrode still contains mercury, this amount of mercury is infinitesimally lower than that in a massive Hg sensor so that the electrode can be used with practically no harm to people or the environment.

We believe that improved accuracy and sensitivity in the bimetallic modified electrodes could be due to the metallic couple's synergistic effects.

Conflicts of Interest

The authors declare no conflict of interest.

Acknowledgment

The authors would like to thank Escuela de Ciencias Químicas y Dirección de Investigación de la Pontificia Universidad Católica del Ecuador (PUCE) and Decanato de Investigación y Desarrollo de la Universidad Simón Bolívar (USB), Caracas-Venezuela for financial support.

References

Abkenar, S.D., Hosseini, M., Dahaghin, Z., Salavati-Niasari, M., & Jamali, M.R. (2010). Speciation of chromium in water samples with homogeneous liquid-liquid extraction and determination by flame atomic absorption spectrometry. Bulletin of the Korean Chemical Society, 31(10) 2813-2818. [ Links ]

Abkenar, S.D., Dahaghin, Z., Sadeghi, H.B., Hosseini, M., & Salavati-Niasari, M. (2011). Determination of zinc in water samples by flame atomic absorption spectrometry after homogeneous liquid-liquid extraction. Journal of Analytical Chemistry, 66(6) 612-617. [ Links ]

Bagheri, A., Behbahani, M., Amini, M.M., Sadeghi, O., Tootoonchi, A., & Dahaghin, Z. (2012). Preconcentration and separation of ultra-trace palladium ion using pyridinefunctionalized magnetic nanoparticles. Microchimica Acta, 178(3-4) 261-268. [ Links ]

Bhatluri, K.K., Manna, M.S., Ghoshal, & A.K., Saha, P. (2017). Separation of cadmium and lead from wastewater using supported liquid membrane integrated with in-situ electrodeposition. Electrochimica Acta, 229 1-7. [ Links ]

Borgo, S. D., Jovanovski, V. Pihlar, & B. Hocevar, S. B. (2015). Operation of bismuth film electrode in more acidic medium. Electrochimica Acta, 155 196-200 [ Links ]

Cargnelutti, D., Tabaldi, L. A., Spanevello, R. M., de Oliveira Jucoski, G., Battisti, V., Redin, M., & Morsch, V. M. (2006). Mercury toxicity induces oxidative stress in growing cucumber seedlings.Chemosphere,65(6), 999-1006. [ Links ]

Dahaghin, Z., Mousavi, H.Z., & Sajjadi, S.M. (2017a). Trace amounts of Cd (II), Cu (II) and Pb (II) ions monitoring using Fe 3 O 4@ graphene oxide nanocomposite modified via 2-mercaptobenzothiazole as a novel and efficient nano sorbent. Journal of Molecular Liquids, 231 386-395. [ Links ]

Dahaghin, Z., Mousavi, & Sajjadi, S.M . (2017b). Synthesis and Application of Magnetic Graphene Oxide Modified with 8‐Hydroxyquinoline for Extraction and Preconcentration of Trace Heavy Metal Ions. Chemistry Select, 2(3) 1282-1289. [ Links ]

Ferreira, M. A., & Barros, A. A. (2002). Determination of As(III) and arsenic(V) in natural waters by cathodic stripping voltammetry at a hanging mercury drop electrode. Analytical Chimica Acta, 459 151-159. [ Links ]

Gumpu, M.B., Sethuraman, S., Krishnan, U.M., & Rayappan, J.B.B. (2015). A review on detection of heavy metal ions in water - an electrochemical approach. Sensors Actuators B Chem., 213 515-533. [ Links ]

Huang, M.-R., Ding, Y.-B., &Li, X.-G. (2014). Combinatorial screening of potentiometric Pb (II) sensors from polysulfoaminoanthraquinone solid ionophore. ACS combinatorial science16 (3) 128-138. [ Links ]

Huang, M.-R., Rao, X.-W., Li, X.-G., & Ding, Y.-B. (2011). Lead ion-selective electrodes based onpolyphenylenediamine as unique solid ionophores. Talanta, 85(3) 1575-1584. [ Links ]

Kefala, G., Economou A., & Voulgaropoulos A. (2004). A study of Nafion-coated bismuth-film electrodes for the determination of trace metals by anodic stripping voltammetry. Analyst, 129 1082 - 1090 [ Links ]

Karri, V., Schuhmacher, M., & Kumar, V. (2016). Heavy metals (Pb, Cd, As and MeHg) as risk factors for cognitive dysfunction: A general review of metal mixture mechanism in brain. Environ. Toxicol. Pharmacol, 48, 203-213. [ Links ]

Lin, J., Zhang, F., & Lei, Y. (2016). Dietary intake and urinary level of cadmium and breast cancer risk: A meta-analysis. Cancer Epidemiol, 42 101-107. [ Links ]

Liang, P., Li, J., & Yang, X. (2005). Cloud point extraction preconcentration of trace cadmium as 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone complex and determination by flame atomic absorption spectrometry. Microchimica Acta, 152(1) 47-51. [ Links ]

Lakshmi, D., Sharma, P. S., & Prasad, B. B. (2007). Imprinted polymer-modified hanging mercury drop electrode for differential pulse cathodic stripping voltammetric analysis of creatine. Biosensors and Bioelectronics, 223302-3308. [ Links ]

Pei, X., Kang, W., Yue, W., Bange, A., Heineman, W. R., & Papautsky, I. (2014). Improving Reproducibility of Lab-on-a-Chip Sensor with Bismuth Working Electrode for Determining Zn in Serum by Anodic Stripping Voltammetry. Journal of the Electrochemical Society, 161 B3160-B3166. [ Links ]

Singh, D.K. & Mishra, S. (2009). Synthesis, characterization and removal of Cd (II) using Cd (II) ion-imprinted polymer. Journal of Hazardous Materials, 164(2) 1547-1551. [ Links ]

Valera, D., Sánchez, M., Domínguez, J. R., Alvarado, J., Espinoza-Montero, P. J., Carrera, P., Bonilla, P., Manciati, C.G. González & Fernández L. (2018). Electrochemical determination of lead in human blood serum and urine by anodic strippingvoltammetry using glassy carbon electrodescovered with Ag-Hg and Ag-Bi bimetallic nanoparticles. Anal. Methods, (10) 4114-4121 [ Links ]

Wan, M. W., Kan, C. C., Rogel, B. D., & Dalida, M. L. P. (2010). Adsorption of copper (II) and lead (II) ions from aqueous solution on chitosan-coated sand. Carbohydrate Polymers, 80, 891-899. [ Links ]

Xiong, S., Ye, S., Hu, X., & Xie, F. (2016). Electrochemical detection of ultra-trace Cu (II) and interaction mechanism analysis between amine-groups functionalized CoFe2O4/reduced graphene oxide composites and metal ion. Electrochimica Acta, 217,24-33. [ Links ]

Zhang, H.F., Shuang, S.M., Wang, G.Z., Guo, Y.J., Tong, X.L., Yang, P. A., Chen, J., Dong, C., & Qin, Y. (2015). TiO2-graphene hybrid nanostructures by atomic layer deposition with enhanced electrochemical performance for Pb(ii) and Cd(ii) detection. RSCAdv., 5 4343-4349. [ Links ]

Zhu, X., X. Zhu, & Wang, B. (2006). Determination of trace cadmium in water samples by graphite furnace atomic absorption spectrometry after cloud point extraction. Microchimica Acta, 154 (1) 95-100. [ Links ]

Zhu, X. S. C., Gao, J. W., Choi, P. L., & Bishop, C. H. (2004). Ahn, On-chip generated mercury microelectrode for heavy metal ion detection. Lab Chip 5 212-217. [ Links ]